CORRESPONDENCE
Year : 2018  |  Volume : 131  |  Issue : 18  |  Page : 2249-2251

Coronary Artery Aneurysm Formation after Drug-Coated Balloon Treatment


1 Department of Cardiology, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Shanghai 200233, China
2 Department of Radiology, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, Shanghai 200233, China

Date of Submission04-Jun-2018
Date of Web Publication10-Sep-2018

Correspondence Address:
Prof. Zhi-Gang Lu
Department of Cardiology, Shanghai Jiao Tong University Affiliated Sixth People's Hospital, No. 600, Yishan Road, Xuhui District, Shanghai 200233
China
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0366-6999.240819

Rights and Permissions

How to cite this article:
Huang J, Zhang JY, Wei M, Yuan MJ, Lu ZG. Coronary Artery Aneurysm Formation after Drug-Coated Balloon Treatment. Chin Med J 2018;131:2249-51

How to cite this URL:
Huang J, Zhang JY, Wei M, Yuan MJ, Lu ZG. Coronary Artery Aneurysm Formation after Drug-Coated Balloon Treatment. Chin Med J [serial online] 2018 [cited 2018 Nov 16];131:2249-51. Available from: http://www.cmj.org/text.asp?2018/131/18/2249/240819



To the Editor: Drug-coated balloon (DCB) has become more and more clinically used as a new interventional treatment technology in recent years. Especially in the treatment of in-stent restenosis (ISR), DCB has demonstrated an unparalleled advantage. However, the related adverse events after DCB treatment are rarely reported. Here, we report two cases of coronary artery aneurysm formation after DCB treatment.

A 62-year-old man suffered from unstable angina received coronary angiography (CAG) in November 2011 which demonstrated that there were diffuse lesions in the proximal segment of the left anterior descending artery (LAD) with complete occlusion of the middle segment. Cypher 2.5 mm × 33 mm and Cypher 2.75 mm × 23 mm stents were implanted sequentially from the middle to the proximal segment of the LAD. These two stents were connected and partially overlapped. Six years later, coronary artery computed tomography (CT) showed a diffuse severe ISR in the LAD with local occlusion and distal stent fracture without artery aneurysm formation [Figure 1]a. Subsequently, CAG confirmed the CT findings, and percutaneous transluminal coronary angioplasty (PTCA) was performed on the LAD. After predilation and postdilation, SeQuent Please (B. Braun Melsungen, Germany) 2.0 mm × 20 mm, 2.5 mm × 30 mm, and 3.0 mm × 30 mm DCBs were sequentially used from the distal to the proximal segment of the LAD dilating for 60 s at 6–10 atm. Retrospective angiography showed thrombolysis in myocardial infarction-3 grade blood flow [Figure 1]b. The patient received CAG again 6 months later, and multiple aneurysmal dilatations could be seen in the LAD [Figure 1]c. Intravascular ultrasound (IVUS) showed that the proximal LAD stent partially fractured while the distal stent completely fractured with multiple stent malappositions and aneurysmal formation [Figure 1]d,[Figure 1]e,[Figure 1]f. Considering that the patient had no symptoms of ischemia and that the distal segment of the LAD had good perfusion, we recommended lifelong dual antiplatelet therapy and close follow-up of the patient.
Figure 1: Imaging appearances of the coronary artery aneurysm formation ([a-f] belong to the first case, [g-j] belong to the second case). (a) CT revealed a diffuse severe ISR in the LAD with local occlusion and distal stent fracture without artery aneurysm formation. (b) CAG showed TIMI-3 grade blood flow after DCB treatment. (c) CAG showed multiple aneurysmal dilatations in the LAD after 6 months. (d-f) IVUS images correspond to the arrows in Figure c, respectively, showed that the proximal LAD stent partially fractured while the distal stent completely fractured with multiple stent malappositions and aneurysmal formation. (g and h) CT myocardial perfusion scintigraphy 1 month after DCB treatment show ed decreased blood perfusion of the left ventricular lateral wall in blue without artery aneurysm formation. (i and j) CT myocardial perfusion scintigraphy 1 year after DCB treatment showed blood perfusion of the left ventricular lateral wall improved, but artery aneurysm could be seen in the proximal segment of the LCx. CT: Computed tomography; ISR: In-stent restenosis; LAD: Left anterior descending artery; CAG: Coronary angiography; TIMI: Thrombolysis in myocardial infarction; DCB: Drug-coated balloon; IVUS: Intravascular ultrasound; LCx: Left circumflex coronary artery.

Click here to view


Another is a 61-year-old man who complained shoulder discomfort after physical activity without the typical ischemic chest pain. Cardiac magnetic resonance presented a perfusion defect in the left ventricular lateral wall from the basal to the middle segment, which was more obvious after adenosine load. CAG revealed that the middle segment of the left circumflex coronary artery (LCx) was occluded with a self-collateral circulation. After predilation, we performed PTCA on the LCx with a SeQuent Please 2.0 mm × 20 mm DCB dilation for 60 s at 6 atm. Retrospective angiography showed a residual stenosis of 30–40% of the middle segment of the LCx. One month after discharge, the patient received a CT myocardial perfusion scintigraphy, which showed that blood perfusion of the left ventricular lateral wall was significantly decreased after adenosine triphosphate load and that the value of myocardial blood flow (MBF) was 52.74 ml/100 ml per minute [Figure 1]g and [Figure 1]h. One year later, we repeated the CT myocardial perfusion scintigraphy. The result was unexpected and showed that the proximal segment of the LCx presented a stenosis of approximately 70% with aneurysm formation and intra-aneurysmal thrombosis following the stenosis. However, blood perfusion of the left ventricular lateral wall improved [Figure 1]i and 1j]. The value of the MBF was 68.23 ml/100 ml/min. Hence, we chose to continue the conservative medical treatment and closely follow up the patient.

The DCB has been increasingly used as a new clinical interventional treatment technology in recent years. Especially in the treatment of ISR, the DCB has demonstrated an unparalleled advantage. However, the related adverse events after DCB treatment are rarely reported. Vassilev et al.[1] first reported the formation of aneurysm two months after DCB treatment for the ISR of the LAD.

Previous perspectives suggested that the possible mechanisms of interventional treatment-related aneurysms include the toxic effects of the drug coatings and local inflammatory and hypersensitivity reactions.[2] Both of the cases we reported used the SeQuent Please DCB. This DCB adopts a patented coating technology called PACCOCATH, wherein the paclitaxel and iopromide mixed matrix is evenly coated on the distal balloon of the traditional PTCA catheter so that the bioavailability of paclitaxel is greatly improved, the contact area of the drug and the arterial wall is increased, the attraction between the drug molecules is weakened, and the cell proliferation and vascular intimal hyperplasia are effectively suppressed. However, in the process of efficiently inhibiting the repair, proliferation, migration, and intimal hyperplasia of vascular smooth muscle and endothelial cells, it will inevitably affect the repair of the intima at the interventional site and delay the healing of blood vessels, which causes thinning of the arterial wall and enlargement of the damaged area in the intima and middle layer, while the pressure in the blood vessel increases, and it will eventually lead to the formation of aneurysms. It is worth noting that DCB has a rapid onset and has a high drug load (3 μg/mm2) than the dosage drug-eluting stent (DES) released (300–600 μg) which undoubtedly amplifies the local toxic effects of paclitaxel. In the first case, multiple DCBs are used in combination which led to higher drug concentration in the overlapping sites and is more likely to cause aneurysm formation. In the second case, we also speculate that the greater local concentrations of the drug are the main reason of aneurysm formation. On the other hand, the operation itself may also lead to aneurysm formation. Too large of the balloon diameter and overexpansion may be the important reason of DCB-related aneurysms, both of which can make the local artery wall teared and thus lead to the formation of coronary aneurysm or even pseudoaneurysms.

There is still much controversy over the treatment of postoperative aneurysm formation and there is also no uniform treatment protocol. The formulation of specific treatment strategies mainly depends on the extent, degree, risk, characteristics of the lesions, and the surgeon's personal experience. Even if aneurysm formation or stent fracture occurred in the two cases we reported, the forward blood flow of the coronary artery was not affected. Hence, no ischemic chest pain and acute coronary events occurred. This may be because DCBs release antiproliferative drugs locally to the coronary arterial wall, which inhibits intimal hyperplasia and reduces inflammatory responses, and as a result, the risk of thrombosis is greatly reduced.[3] According to the above reasons, we only chose dual antiplatelet therapy and followed the patient closely long term.

In summary, we propose the potential risk of aneurysm formation after DCB treatment through these two cases, which might be associated with the high dose, high concentration, and rapid onset of cytotoxic drugs coated on DCB. The safety of DCB should be further consideration and discussion.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form, the patients have given their consent for their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identities, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Vassilev D, Hazan M, Dean L. Aneurysm formation after drug-eluting balloon treatment of drug-eluting in-stent restenosis: First case report. Catheter Cardiovasc Interv 2012;80:1223-6. doi: 10.1002/ccd.23460.  Back to cited text no. 1
    
2.
Singh H, Singh C, Aggarwal N, Dugal JS, Kumar A, Luthra M. Mycotic aneurysm of left anterior descending artery after sirolimus-eluting stent implantation: A case report. Catheter Cardiovasc Interv 2005;65:282-5. doi: 10.1002/ccd.20338.  Back to cited text no. 2
    
3.
Cremers B, Toner JL, Schwartz LB, von Oepen R, Speck U, Kaufels N, et al. Inhibition of neointimal hyperplasia with a novel zotarolimus coated balloon catheter. Clin Res Cardiol 2012;101:469-76. doi: 10.1007/s00392-012-0415-7.  Back to cited text no. 3
    


    Figures

  [Figure 1]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
References
Article Figures

 Article Access Statistics
    Viewed163    
    Printed1    
    Emailed0    
    PDF Downloaded44    
    Comments [Add]    

Recommend this journal




京ICP备05052599号